Products


3M1264  Indicator Attest Eth Oxide 100/bx 4Bx/cs 3M

Product: 3M1264 Manufacturer: 3M Healthcare Stock Item? No

Please sign in for pricing.

Please Wait...
  • ETHYLENE OXIDE INDICATOR - SEALED GLASS AMPULE. Biological indicators provide the best assurance of sterility by challenging the sterilizer with quantifiable, highly resistant spores. The system includes biological indicators, disposable test packs, incubators and log books. Biological indicators are used for load control monitoring which is the process by which a load is monitored and released based on the result of a biological indicator (BI) in a test pack. 3M ATTEST 1264 OR 1264P BILOLOGICAL INDICATORS FOR ETHYLENE OXIDE: The self-contained (BI) for all EO cycles consists of a Bacillus subtilis spore strip; sealed glass ampule with growth medium and bromothymol blue pH indicator dye system; green color-coded cap with holes for sterilant penetration and a Tyvekr filter as a bacterial barrier, and a chemical indicator on the label that changes to green when processed. After sterilization, the vial is "crushed" to join the growth media with the processed spore strip. The BI is incubated for 48 hours for a visual color change readout. A color change to yellow indicates surviving spores and a positive result. 3M ATTEST 1278 OR 1278F ETHYLENE OXIDE PACK: A pre-assembled, EO plastic disposable test pack with an Attest 1264 biological indicator at the end of a 50,8cm or 20 inch tortuous or winding path. The pack is used to monitor all EO cycles. The pack is designed to present a resistant challenge to the EO sterilization process as is the AMMI routine syringe test pack.

Attention CA customers:

California propostion 65 Warning label